37
For editing, select the field you want to change and click
Edit
.
To delete the unnecessary field press the
Delete
button
.
If you have accidentally damaged the plate and can not carry a calibration procedure, you need to order a
new one from the manufacturer.
Calibration for scanning zone
№
0
To obtain a three-dimensional model of large objects, RаngeVisiоn 3D scanner can be optionally equipped
with a scanning zone №0 with the scanning area 850x530x530 mm.
This scan area can be used only if there is already a scanning zone №1. To configure the scanning zone
№0:
Install lenses with focal length of 9 mm and Blue light filters on the cameras.
Perform calibration for the scanning zone №1, by specifying used 9mm lenses in the menu of full
calibration . Working distance to the field will be about 70 cm.
By loosening the By loosening the mounting, extend camera at the maximum possible distance
(about 35 cm).
Install calibration plate for the zone №1 at a distance of 105cm from the cameras.
Point the cameras at a single point, switching the Cross lighting mode and tighten the mounting.
Perform
Orienting