CP-08-01
9040 Classic HV Power Supply - Maintenance
2 0
2 0
2 0
2 0
2 0
Figure 6: Current Output Test
Figure 6: Current Output Test
Figure 6: Current Output Test
Figure 6: Current Output Test
Figure 6: Current Output Test
Bench T
Bench T
Bench T
Bench T
Bench Testing
esting
esting
esting
esting
Equipment Required:
Equipment Required:
Equipment Required:
Equipment Required:
Equipment Required:
- Volt / Ohmmeter
The "Troubleshooting Guide" provides information
for troubleshooting the power supply when improper
operation is obtained and the problem has been
traced to the power supply. Proper troubleshooting
should ONLY be accomplished with specific test
equipment by qualified electronics technicians or
authorized ITW Ransburg representatives.
Before troubleshooting, ensure that the power
supply is plugged into a live outlet of the appropriate
voltage. All electrical measurements in the
"Troubleshooting Guide" are nominal and may
vary as much as
+
/- 10% depending on the test
conditions and the test equipment used. Refer to
Figures 6 and 9 for location of parts called out in the
"Troubleshooting Guide" in this section.
For bench testing, jumper JP4 should be used to
trigger the high voltage output (see "PC Board
Test Jumper" discussed at the beginning of the
"Maintenance" section).
>
The high voltage multiplier and high
voltage cable contain significant capaci-
tance that will store charge. ALLOW
ALLOW
ALLOW
ALLOW
ALLOW
approximately 10 seconds for this charge
to bleed off BEFORE
BEFORE
BEFORE
BEFORE
BEFORE opening the cabinet
door OR REMOVING
OR REMOVING
OR REMOVING
OR REMOVING
OR REMOVING the high voltage
cable from the power supply or spray
applicator.
C A U T I O N
C A U T I O N
C A U T I O N
C A U T I O N
C A U T I O N
!!!!!
>
The troubleshooting procedure outlined
in the "Troubleshooting Guide" requires
measurement of voltage potentials that can
cause SERIOUS BODILY INJURY
SERIOUS BODILY INJURY
SERIOUS BODILY INJURY
SERIOUS BODILY INJURY
SERIOUS BODILY INJURY if
proper measuring procedures are not
followed. For this reason, proper trouble-
shooting should ONLY
ONLY
ONLY
ONLY
ONLY be accomplished
with specific test equipment by qualified
electronics technicians or authorized ITW
Ransburg representatives.
W A R N I N G
W A R N I N G
W A R N I N G
W A R N I N G
W A R N I N G
!!!!!
>
Before going through the trouble of a
complete bench test, jumper JP4 on the
main PC board to trigger the high voltage
and see if the suspected problem goes
away. If it does, replace the high voltage
trigger device (flow switch, pressure
switch, PC contact, etc.) or repair the in-
ternal wiring of the power supply con-
nected to the trigger device, and test
again. If the problem does not go away,
then proceed with a bench test.
N O T E
N O T E
N O T E
N O T E
N O T E
ARCHIVE