ratiotec GmbH & Co. KG | Tel.: +49 201 361 490 | Fax: +49 201 361 49 26 | Email: info@ratio-tec.de
2.
1.
3.
Press the
BATCH
button and keep it
pressed as you turn on the device. The
device switches to the calibration mode.
The following display appears. Release the
BATCH
button.
Place the white calibration paper in the
banknote feeder or banknote ejector, so
that the sensors are completely covered
(Illustrations 3). Then press the
BATCH
button. Calibration starts automatically.
The calibration mode may take a few sec-
onds. The following display appears.
To finish, press the
RESET
button and then
briefly the
START
button. The white cali-
bration paper is ejected immediately from
the banknote ejector. The device com-
pletes the calibration process. You may
turn off the device and start again.
1-2625
F9F9
1-2624
1E1B
CALIBRATION GUIDE rapidcount S-Serie
[EN]
Calibration Guide with calibration paper:
ratiotec GmbH & Co. KG | Tel.: +49 201 361 490 | Fax: +49 201 361 49 26 | Email: info@ratio-tec.de
white
calibration paper
white
calibration paper
REMARK: the calibration manual includes illustrations as examples, the values of the
sensors may differ.