Instructions for Use MPR1 DATALOGGER
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
Page 26 of 98
3.4.2 Cleaning, disinfecting and sterilising the temperature sensor
RAUMEDIC multi-parameter catheters (precision pressure catheters with additional temperature
measurement) are sterile products and are intended for single use only.
3.5 Analogue outputs
Type:
resistance output, ratiometric characteristics
Where:
Figure 2: Connection field, connections [21] and [25]
Application: signal transfer
The MPR1 has two identical analogue outputs to transfer up to two invasively measured pressure signals
to a third-party system (bed-side monitor) with alarm function.
Use approved accessories only.
Every person who connects an additional unit to the signal input or signal output configures a medical
system, and thus he / she shall be responsible for the fact that this system meets the requirements of
standard IEC 60601-1.
Please keep in mind that the errors in the pressure display multiply on the third-party system when
analogue outputs are used. Error of the pressure error of the pressure amplifier of the MPR1
+ error of the analogue error of the pressure amplifier of the third-party system. When using
the analogue output, it is imperative that you proceed in keeping with chapter 4.6.1
As from 17th March 2004 the following third-party systems shall be considered approved:
•
Armeda PM 9000
•
Marquette Dash 3000
•
Marquette Hellige SMU 611
•
Marquette Hellige EAGLE 4000
•
Hewlett Packard M1166 Model 66S
Please enquire with the manufacturer for other third-party units.
Cable selection:
depending on the apparatus, cf.
If you want to transfer the signals to another third-party system not listed here, please get in touch with the
manufacturer. You will receive updated information on the cable systems available and the third-party
systems approved.
3.6 Interface for the extension of the apparatus
Type:
RS232
Number:
1
Where:
Figure 2: Connection field, connection [22]
Application:
unit service and optional extension of the apparatus (preparation
only)
This interface is used for device (apparatus) service, diagnostics and maintenance as well as for updating
the software. Alternatively, optional device accessories from RAUMEDIC can be connected to this interface.
Currently the following RAUMEDIC accessories are available for connection to this interface: reader TDT1
readP for telemetric ICP measurement. (see
Telemetric pressure measurement)