Instructions for Use MPR1 DATALOGGER
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
Page 52 of 98
4.4.4.3 Temperature trend
In order to skip from the graphics screen to the trend screen , press the key . The navigation field
renders available keys to set the temporal resolution as well as page up and page down keys. Two different
speeds are available for paging up and down. If you keep one of the navigation keys pressed for a longer
period of time, the key function is carried out again through an automatic routine. Thus, you can get an
overview of the entire trend fast without having to press the key several times; subsequently you can set a
higher resolution for the time range of your choice.
Six time ranges are available (width of the time axis in the trend graphics) which you can switch over by
using the key
and
.
Finer temporal resolution
Coarser temporal resolution
Rough backward navigation
Fine backward navigation
Fine forward navigation
Rough forward navigation
4.4.5 Telemetric pressure measurement
Observe section 3.3 Telemetric measurement of the ICP.
Use only approved accessories
Observe the instruction manuals of the accessories
Adhere to the permissible measurement range of the transducer
The MPR1 has one input socket (black colour coding) for connection of accessories (reader TDT1 readP).
Socket for the connection of the TDT1 readP
If the reader TDT1 readP is connected to this socket, then it is automatically recognized by the MPR1 and
the search run for a telemetry catheter is started. The time of the search run is indicated by the timer in the
ICPT table field (see 6.14 Reaction to fault messages and indications). To make the connection to a
telemetry catheter, the reader aerial must be placed close to a telemetry catheter. If a telemetry connection
is established to a telemetry catheter, this is automatically assigned to the ICPT measuring point.