5 CALIBRATION
CALIBRATION INSTRUCTIONS INDICATOR 110
DETERMINE THE SOFTWARE VERSION
Switch the system on.
❏
The indicator shows the following sequence:
8888.8 (testing LCD segments)
7.x (software version)
13425 (the calibration number)
(current weight, standard weigh mode).
DEFINING ZERO
Unload the system.
Switch the system on.
Push the
0/T
key for about 8 seconds.
❏
The display counts down from AF 08 to AF 00.
❏
The indicator shows which percentage of the total capacity has been zeroed, e.g. AP 6.4. This
percentage should not be higher than 20.
❏
The zero point has now been defined, the system automatically returns to standard weighing
mode.
CALIBRATION
The indicator offers the possibility to enter a maximum of three calibration points (multiple point
calibration). The advantage is that even weighing systems with bad hysteresis can be calibrated within
specifications.
Since these instructions are often used in the field, where it is difficult to calibrate various points, we
will start with the explanation of single point calibration.
SINGLE POINT CALIBRATION
Deletion (resetting to zero) of earlier calibrated points
Push the
PT key for about 18 seconds (this may be 8 seconds on older versions)
❏
The display will go blank until calibration mode has been reached.
❏
The indicator shows the value of the
first
calibration point, the
lowest
indication bar (on the left
of the display) is flashing.
Use the
and
keys to see the three earlier programmed values on the screen by moving the
indication bar up and down.
❏
When the lower indication bar is lit, the first (lowest) value is shown
❏
When the higher indication bar is lit, the second (middle) value is shown.
❏
When both indication bars are lit, the third (highest) value is shown
.
❏
When calibrating only one point the second and highest values should be returned to zero.
Press the
or
key until the middle calibration value is in the screen.
❏
The higher indication bar is flashing.
Press the
0/T
key.
❏
The first segment is flashing.
Use the
and
keys to set the flashing segment to zero.
Change to the next segment by pressing the
0/T
key.
Set all the segments to zero until the indication bar is flashing.
Press the
or
key until the highest calibration value is in the screen.
❏
Both indication bars are flashing.
Repeat the process until all the segments are set to zero.
Summary of Contents for 100 Series
Page 7: ...7 3 1 Accurate Weighing 2 ...
Page 8: ...8 MIN 3 2 Touch Panel Indicator 3 3 Net Tare Gross 1 2 3 ...
Page 9: ...9 3 4 On Off ...
Page 10: ...10 3 5 Error Messages max ...
Page 11: ...11 ...
Page 12: ...12 3 6 Zero and Tare Functions ...
Page 13: ...13 3 7 Manual Tare Entry ...
Page 14: ...14 3 8 Summing ...
Page 15: ...15 3 9 Total and Reset ...
Page 22: ...22 ...
Page 23: ...23 ...