ERROR MESSAGES INDICATOR
Displayed error Meaning
Out of error mode
Err01
Load cell signal is unstable
Disappears when signal is
stable again
Err02
Overload on full weighing system
Disappears when overload is
removed
Err03
Gross negative. This action is not allowed
Disappears after 3 sec. No
Tare action will be accepted
Err04
Out of zero range
Press any key
Err05
Sampling accuracy too low
Press any key
Err06
Input signal too high (over load positive)
Check weighing system
(load cells + cabling)
Err07
Input signal too low (over load negative)
Check weighing system
(load cells + cabling)
Err08
Calibration out of range (negative)
Follow correct calibration
procedure
Err09
Calibration out of range (signal too low)
Follow correct calibration
procedure
Err10
Calibration count 2
nd
(3
rd
) point lower than
count 1
st
(2
nd
) point
Follow correct calibration
procedure
Err11
Calibration from within piece counting mode
Follow correct calibration
procedure
Err14
Setpoint value 2 < setpoint value 1. This is not
allowed
Enter limit values correctly
Err20
Err97
Signal of load cell is not correct during start-up
of indicator
Calibration locked (jumper JP1 placed)
Check weighing system
Place Jumper JP1. Located
next to the –EX connection
Err98
Calibration point must be higher than previous
one
Follow correct calibration
procedure
Err99
-----
Or
-----
Action only allowed in start-up units (kg/lb)
Lower than – 2% of full scale value
Level ERROR > 2% non-level
Press ON/OFF (CE) key
Check weighing system
(mechanical) perhaps even
execute a zero calibration
Place weighing system in
horizontal position