USER MANUAL
– Safecheck
28
9.4
The touch panel of the Safecheck
Each key has an operational and an entry function.
Operational function
Entry function
zero setting and
automatic tare
confirm and
digit to the left
enter set point
(limit value)
decrease flashing digit
totaling
on/off
increase flashing digit
9.5
Enter setpoint value (limit value)
IMPORTANT:
To activate the limit value of the relays you have to go into the parameter menu and
change P13 to ‘0’. Go back into the weighing mode and enter the limit value by pressing the
PT key.
A setpoint can be used to activate a buzzer.
To activate the last entered setpoint value:
Press the
PT key.
The display shows the current setpoint value.
The digit on the right flashes.
Press ENTER (
↵
) for three seconds if the current setpoint value is required.
To enter a new setpoint value:
Press the
PT key.
The display shows the current setpoint value.
The digit on the right flashes.
Press the
∧
key to go up a value or press the
∨
key to go down a value until the required value is
reached.
Press ENTER (
↵
) to change the next value.
Repeat this procedure until the required limit value is displayed.
Press ENTER (
↵
) until the last segment has stopped flashing.
The new setpoint is saved and activated.
The gross weight is being displayed.
The new setpoint will be active until a new setpoint value has been entered.
NOTE: After the limit value is set, put P13 back to ‘1’!
The relay output will be activated when the gross weight exceeds the setpoint value and will be
deactivated subsequently when the gross weight becomes less than the setpoint value.