32
power, Power supply, moisture condensation is abnormal, the laser will send out alarm
information.
If any alarm occurs (except for Interlock alarm) when the laser is running, the
homepagewill display the alarm that occurs, and the ALARM light (yellow) on the front panel of
the laserwill light up, the laser will stop emitting and lock.
When the Interlock is abnormal, the output of Ready signal is low, and the Interlock state
isabnormal in the software homepage, but the laser is not locked, and the ALARM light
(yellow)is not lit. The output of Ready signal is on high level when the Interlock is normal.
5.2 Alarm Solutions
The instructions and solutions of alarms are as follows:
Table 12 Instructions and Solutions for Alarms of Laser
Alarm name
Alarm instructions and solutions
System Timer Alarm
Instruction:
The internal clock of laser is abnormal.
Solution:
If it displays this alarm, please contact Raycus.
Laser On Button Alarm
Instruction:
This alarm occurs, The ‘LASER’ button on the front panel has been pressed
before the laser is powered on.
Solution:
Please power off and release the “LASER” button. Repower and check if the
alarm is resolve. If this alarm continues to occur, please contact Raycus.
InterLock Alarm
Instruction:
Occurs when the InterLock is disconnected.
Solution:
Short-circuit the interLock pins and restart the laser to try. If this error continues
to occur, please contact Raycus.
Current Driver Alarm
Instruction:
Current Driver Alarm, this error occurs when the constant current driver board
inside the laser is abnormal.
Solution:
Restart the laser, If this error continues to occur, please contact Raycus.
Laser Power Alarm
Instruction:
The laser will get this error when it does not emit with correct settings. Power
error occurs only when the laser is emitting.
Solution:
Restart the laser, if this error continues to occur, please contact Raycus.
ACDC1 Alarm
Instruction:
ACDC1 Error, failure of the laser power supply or sudden power failure of the
power supply system may cause an alarm.
Solution:
Check if the input AC voltage is normal. Restart the laser, if this error continues