Remind you to change the bacteriologic filter. An acoustic signal will sound intermittently.
To delete this message it must be stopped and started with the main switch (3).
WATER IN CHAMBER
The control detects the water inside the sterilization chamber, out of cycle. This message
will be alternative with another: “START TO CONTINUE”, that shows the possibility to re-
start the cycle again. This situation will take place after the execution of a program of
liquids or when the cycle has been aborted in a step where the pressure was not enough
to take out the water in a solid program.
It can also be a failure. (See FAILURE MESSAGES)
FAILURE MESSAGES
The failures detected during the sterilization process, a part from giving you the printing
warning, are accompanied with the sound of an acoustic signal and the light of LCD.
Pressing the switch “START/STOP”, the sound stops. When pressing the switch will
disappear the failure warning message also, and the lights of the LCD., in case you don’t
have to do a discharge phase (program LIQUIDS), if the cycle was in a determined phase,
the control still consider to be pressure inside the autoclave, or the temperature is still
high, in the case of liquid programs when the cooling process is finished or the discharge,
the acoustic signal will sound.
FAULT PROBE 1
The control detects that the temperature probe 1 (main) is broken. It is impossible to
proceed with the sterilization. If during the sterilization the probe is broken the process
stops.
FAULT PROBE 2
The control detects that the probe 2 (heart probe) is broken during a sterilization process
with two probes. If this happens during a cycle, the sterilization is aborted.
In this case it can be continued the sterilization while the probe 2 is changed, by editing
the programs as SOL-1 and LIQ-1.