In program mode it is used to decrease the value of the parameter in edition.
Out of program it is also used to select a program.
11.- Programming switch. (PRO)
While pressing this switch during 2 seconds you can access to the edition of the program.
During the process of sterilization will shoe the program while pressing.
12.- Start/Stop switch.
This switch is used to run or to stop the sterilization program selected.
To stop the sterilization cycle it must be pressed the START/STOP switch during 2
seconds. If this operation is made, in the final onform will appear the message “STOP”
instead of the message “STERILIZATION OK”.
TECHNICAL SPECIFICATION
Apparatus:
Steam
sterilization autoclave
Sterilization
temperature: From 105ºC to 139ºC.
Max. pressure: 2,5 bar
Voltage: 220 V single-phase (for AE-28 DRY and AE-75 DRY)
220 V or 380 V triple phase + N (for AE-110 DRY and
AE-150 DRY)
Frequency:
50/60
Hz
Consume: 2.000 W (Model AE-28 DRY)
3.000 W (Model AE-75 DRY)
4.500 W (Model AE-110 DRY)
6.000 W (Model AE-150 DRY)
Sterilization time: from 3 to 120 minutes.
Final Drying: Selectable: (DRY/NOT DRY)
Drying time: From 0/3 to 60 minutes.
Data
printing:
Manual.
Printing cadence: Every 120 seconds.
Printer connection: Interface RS-232