Preliminary
4: Thermal Printer
Allows you to quickly get your results on paper by using the built-in thermal printer.
5: Safety Indicator
Green Light on: System is in safe mode. Ready for handling test leads, connections.
Red Light on:
Unsafe mode! Don’t remove any cables and don't touch the test object.
6: Display with Touch Panel
The instrument can be easily manipulated and fully controlled by the touch screen panel. If
preferred, an external mouse or keyboard can be connected to the USB Port to control the system.
7: Safety GND
Connect a proper earth.
The FIRST cable you have to connect and lead to earth ground is always the Safety GND. This is
also the LAST cable to disconnect! Any interruption of the grounding connection can create an
electric shock hazard.
8: Measurement Input B
Connect the red measurement cable to this input.
9: Measurement Input A
Connect the blue measurement cable to this input.
10: HV GND
This is also a measurement Input. Connect the green-yellow cable to this input.
11: HV Output
This is the receptacle for the high voltage. Connect the orange high voltage cable to this input.
1
This is not the same plug as the one from the CAPO 12, the cables are not interchangeable!
90315-1.2 Instruction Manual CAPO
Page 34 of 117
WARNING !
WARNING !
High voltage is on!