Accessories
MP150
39
Note
: The fittings needed to connect the inner taper thread of the linescanner tube fitting (Rc
1/8") and the cooling hose of the Thermostat are not supplied as standard components! We
recommend a
® Stainless Reducing Union (8 mm outer diameter – 6 mm inner
diameter) and
•
Move the external temperature sensor into the mounting block and fix it using the M3x8 hex
screws Take care not to damage the sensor when tightening the screws! Do not bend the
connection line between sensitive element and thermostat at a radius tighter than 5 mm!
•
Adjust the thermostat according to the table below.
Relative Humidity [%]
Am
bi
ent
Te
m
pe
ra
tur
e [
°C
/°F]
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95 100
0/32
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
5/41
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
10/50
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
15/59
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
0
20/68
+
+
+
+
+
+
+
+
+
+
+
+
+
0
0
0
0
0
1
25/77
+
+
+
+
+
+
+
+
+
+
0
0
0
1
1
1
1
1
2
30/86
+
+
+
+
+
+
+
0
0
0
1
1
1
1
2
2
2
2
3
35/95
+
+
+
+
+
0
0
1
1
1
2
2
2
2
3
3
3
3
4
40/104
+
+
+
+
0
1
1
1
2
2
2
3
3
3
4
4
4
4
5
45/113
+
+
0
0
1
2
2
2
3
3
4
4
4
4
5
5
5
5
6
50/122
+
+
0
1
2
2
3
3
4
4
4
5
5
5
6
6
6
6
7
60/140
0
1
2
3
3
4
5
5
5
6
6
7
7
7
7
7
7
7
–
70/158
1
2
4
4
5
6
6
7
7
7
7
7
–
–
–
–
–
–
–
80/176
2
4
5
6
7
7
7
–
–
–
–
–
–
–
–
–
–
–
–
90/194
4
5
7
7
7
–
–
–
–
–
–
–
–
–
–
–
–
–
–
100/212
5
7
7
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
+
thermostat not necessary
0
thermostat recommended
1 – 7 thermostat necessary, 1 – 7: recommended adjustment of the thermostat
–
housing temperature out of specification
Tab. 4: Thermostat Adjustment
•
Check the system functionality. After a few minutes the internal housing temperature should
reach the minimum device temperature as follows:
Thermostat adjustment Minimum device temperature
0
15°C / 59°F
1
20°C / 68°F
2
25°C / 77°F
3
30°C / 86°F
4
35°C / 95°F
5
40°C / 104°F
6
45°C / 113°F
7
50°C / 122°F