RBR#0008818revB
24
1.
2.
3.
4.
5.
6.
7.
1.
2.
3.
1.
2.
Step
Description
3
Fill the buffer tube with oil
Remove the stopper from the syringe
Install the needle
Invert the syringe so that the needle is point up
Purge the air from the needle by depressing the plunger until a drop of
oil comes out
Insert the needle into the buffer tube assembly and fill it with oil to the
top of the set screw
When extracting the needle, continue to apply pressure to the plunger
to maintain the oil level
Draw the oil out of the needle and remove the needle
4
Refilling the oil
Install the refill adaptor to the syringe
Invert the syringe so that the refill adaptor is pointing up
Purge the air from the refill adaptor by depressing the plunger until a
drop of oil sits at the adaptor opening
5
Install the refill adaptor
Mate the refill adaptor to the pressure port
Install the four cap screws with the 5mm hex key
Ideally, the meniscus at the air-oil interface should be convex to
minimise the air in the final assembly.
Do not apply pressure to the plunger when the syringe is installed
on the pressure port! Doing so may exceed the pressure rating of
the sensor.