background image

3

Minimum Installation Requirements

Minimum Requirements

• 

Input Voltage:  

100 ~ 120VAC ~ 60Hz, 48 W

• 

Output Voltage:  

24 ± 0.5VDC / 2A.

 

• 

Standby power consumption:  

≤0.5 W

• 

DC ripple:  

100mV (P-P)

• 

Ambient Temperature:  

-10 ~ + 50 °c

• 

Large, full metal casing provides excellent heat dissipation to protect the electrical power 
components.

• 

Overload and output short circuit protection

• 

Frequency bandwidth: 

5 ~ 1000MHz

• 

Throughput loss:  

≤0.5 dB

• 

Band flatness:  

± 0.5dB

• 

Return loss:   

≥ 13dB

• 

Input / Output impedance:  

75 Ω

ADDITIONAL NOTES:

Contact RCA Commercial for a complete list of compatible arm models that can be used with 

this wall bracket power supply.

The JPS24D is an efficient switching power supply that provides DC24V / 1 A supply output 
combined with the TV input signal on a COAX output. This product is used in hospitals, hotels, 
RVs, boats, and other television signal transmission systems incorporating the RCA J14HV840 
television designed to be powered over the COAX cable.

At the time of this printing, these configurations have been evaluated and tested by Under-

writers Laboratories. Other means of mounting may be potentially hazardous.

ARM MODEL NOTES:

SPECIFICATIONS

Overview & Cleaning

CLEANING :

  Clean the exterior of this mount by removing dust with a lint-free cloth. For further cleaning, 

use a soft cloth or paper towel dampened with water.

CAUTION :

  To avoid damage to the surface of the mount, do not use abrasive or chemical cleaning agents.

DISINFECTING :

  Do not immerse this mount, rather clean with a soft damp cloth. To avoid damage to the 

surface of the mount, test a small portion of the mount’s cabinet with any new disinfectant to verify that the 

disinfectant will not discolor or soften the enclosure.

This apparatus with class I construction shall be connected to a main socket outlet with a protective earthing

connection.

CLEANING & DISINFECTION:

Summary of Contents for JPS24D

Page 1: ...Changing Entertainment Again LED Television User s Guide Hospital Grade Television Mount Installation Guide JPS24D ...

Page 2: ...ng this product allows us to contact you if needed Keep your sales receipt to obtain warranty parts and service and for proof of purchase Attach it here and record the serial and model numbers in case you need them These numbers are located on the product TO PREVENT FIRE OR SHOCK HAZARDS DO NOT EXPOSETHIS PRODUCTTO RAIN OR MOISTURE This reminder is provided to call the CATV system installer s atte...

Page 3: ...her television signal transmission systems incorporating the RCA J14HV840 television designed to be powered over the COAX cable At the time of this printing these configurations have been evaluated and tested by Under writers Laboratories Other means of mounting may be potentially hazardous ARM MODEL NOTES SPECIFICATIONS Overview Cleaning CLEANING Clean the exterior of this mount by removing dust ...

Page 4: ...be able to support a minimum loading of 350 PSI applied perpendicular to the mounting hole locations by compression or tension Some wall construction may require internal bracing to accept the minimum loading required 3 HARDWARE REQUIREMENTS Mounting fasteners are NOT supplied with the wall bracket 4 TOTAL MAXIMUM WEIGHT of the TV receiver and support arm system cannot exceed 41 lbs 1 All threaded...

Page 5: ...e side of the patient s bed Please note also that the suspension arm requires a min imum clearance distance of 30 from ceiling to the top of the wall bracket A check with the contractor architect or Facility Engineering Department regarding hidden objects behind the wall should be made before proceeding with this installation and final operation of the arm assembly 2 Level the wall mount pivot hol...

Page 6: ...TO RAIN OR MOISTURE CAUTION TO REDUCE THE RISK OF ELECTRIC SHOCK DO NOT REMOVE COVER NO USER SERVICEABLE PARTS INSIDE REFER SERVICING TO QUALIFIED SERVICE PERSONNEL CAUTION MAIN PLUG OF THE POWER SUPPLY UNIT IS THE METHOD OF DISCONNECTING POWER THE PLUG MUST REMAIN READILY ASSESSABLE CAUTION ASSESSABLE METAL PARTS OF THIS APPARATUS ARE ELECTRICALLY ISOLATED FROM THE GROUNDING CONDUCTOR OF THE SUPP...

Page 7: ...le specified by the manufacturer or sold with the apparatus When a cart is used use caution when moving the cart apparatus combination to avoid injury from tip over IMPORTANT SAFETY INSTRUCTIONS Read before operating equipment 13 Unplug this apparatus during lightning storms or when unused for long periods of time 14 Refer all servicing to qualified service personnel Servicing is required when the...

Page 8: ...tv com DTI Services LLC 5350 Lakeview Parkway Dr South Indianapolis IN 46268 1 800 722 2161 DTI Please do not send any products to the Indianapolis address listed in this manual or on the carton This will only add delays in service for your product ...