92
VISUAL INSPECTION
Are the legs or arms
fastened properly during
leg or arm/-upper body
training?
Are any other electrical
devices, unauthorized as
medical products, in the
patient's reach?
Execute the fastening of the
legs/arms as indicated in the
last preceding pages.
Please remove all other
electronic devices, not
authorized as medical
products, from user's
(patient's) reach.
During the course of training,
legs and arms should not move
unwontedly out of the foot shells
or the forearm shells.
Should the patient come in
contact with a »low quality«
mains-supplied device during
the course of training, the user
protection from an electric
shock is is no longer guaranteed
by the high standard isolation
of the MOTOmed device but is
determined by the properties of
the »low quality device«. In order
to effectively provide full medical
product safety standard of the
MOTOmed, all other unauthorized
mains-powered medical devices
are to be removed from the user's
reach.
Test step
Necessary measures in
case of fault detection
during visual inspection
Explanation