RecoveryPump
®
Model 737R (RPX)
User Manual
c. Make sure the air hoses are not bent, kinked, or pinched.
Note:
When only one garment is used, the second outlet must be plugged with
the Prong Plug. Otherwise the console will stop working and the alarm
buzzer will beep.
Step 6: Select the Treatment Program
a. Turn On the Main Switch.
The following screen is displayed, showing the device name and model,
and the Firmware version.
After two seconds, the following screen is displayed, showing the last
selected program and its pressure and time settings.
b. Press the
Mode
button.
c. Press the
Up/Down
button to select the required program.
Step 7: Set Treatment Pressure
a. Press the
Mode
button.
The
Treatment Pressure
screen is displayed.
b. Press the
Up/Down Button
to select the treatment pressure.
Note:
The minimum pressure setting is 20 mmHg. The maximum pressure
setting is 100 mmHg.
20
L98005H1R-A