RecoveryPump
®
Model 737R (RPX)
User Manual
2. RecoveryPump
®
Specifications and Dimensions
Model: 737R, appliance coupler, transportable, rated: 100-240 VAC, 50-60 Hz
Type of protection against electric shock
Class II
Degree of protection against electric shock
Type BF
Degree of protection against ingress of water IP21
Degree of safety of application in the
presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide
Equipment not suitable for use in the
presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide
Mode of operation
Continuous
Treatment duration
15 minutes to 90 minutes
Environmental Conditions
Normal: 10-30°C, 30-75% rH, 700-1060 hPa
Pressure Range
20 mmHg to 100 mmHg
Size
22 cm × 17 cm × 13 cm (8.6" × 6.7" × 5.1")
Weight
1.5 kg (3.3 lbs.)
3. Labels
The following labels and symbols appear on the console, garments and/or packaging.
Symbol
Description
Location
IP21
Degree of protection against
ingress of water
On base of console
Read instructions before use
On base of console
Read instructions before use
On garment label
Level of protection type BF
equipment
On base of console
Double insulation
On DC power adapter
Direct current
On base of console
Manufacturer's name and address On garment label & on base of
console
Date of Manufacture
On base of console
L98005H1R-A