background image

RP Lite

®

 Model 760R 

RP Sports 

User Manual

11. EMC Manufacturer Declarations 

Electromagnetic Compatibility (EMC) Statement for Home Healthcare 
Environment 
The RP Lite

®

 system has been evaluated to international standard 

IEC 60601-1-2 "General requirements for basic safety and essential 
performance - Collateral standard: Electromagnetic compatibility - 
Requirements and tests". 
The Medical Electrical Equipment needs special precautions regarding EMC 
and needs to be installed and put into service according to the EMC information 
provided here within the accompanying documents. 
Portable and mobile RF communication can affect the Medical Electrical 
Equipment. See below recommended separation distances between portable 
and mobile RF communication equipment and the RP Lite

®

 system. 

Wireless communications equipment such as wireless home network devices, 
mobile phones, cordless telephones and their base stations,  
walkie-talkies can affect this equipment and should be kept at least a distance  
d = 3.3 m away from the system. 

*Note

: The distance calculated from 800 MHz to 2.5 GHz 

Rated Maximum Output 

Power of Transmitter (W)

Separation Distance* according to 

Frequency of Transmitter (m)

0.01

0.23

0.1

0.73

1

2.3

10

7.3

100

23

Caution: 

Do not use a mobile phone or other devices that emit electromagnetic 
fields, near the unit. This may result in incorrect operation of the unit. 

This system should not be used adjacent to or stacked with other 
equipment and that if adjacent or stacked use is necessary, this 
machine should be observed to verify normal operation in the 
configuration in which it will be used.

Model 760R - Electromagnetic Emissions – Manufacturer’s Declaration

The 

Model 760R 

is intended for use in the electromagnetic environment specified below. The 

customer or the user of the 

Model 760R 

should assure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF emissions 

CISPR 11

Group 1

The 

Model 760R 

uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not likely to cause 

any interference in nearby electronic equipment.

 

L98107R0-G

36

Summary of Contents for RP Lite Model 760R

Page 1: ...RP Lite Model 760R Massage System User Manual RP Sports 9 Lacrue St Ste 108 Glen Mills PA 19342 855 732 7867 www rpsports com L98107R0 G...

Page 2: ......

Page 3: ...te Specifications and Dimensions 3 3 Labels 4 4 Indications for Use 7 5 About RP Lite 7 6 Components of the RP Lite System 9 7 How to Operate the RP Lite System 14 8 Maintenance and Storage 28 9 Troub...

Page 4: ......

Page 5: ...ing Explosion Hazard Do not use the RP Lite system in the presence of flammable gases including flammable anesthetics Electric Shock Hazard Do not allow liquid to enter any part of the RP Lite console...

Page 6: ...stem keep it away from dust lint and dirt Keep away from sources of heat or moisture Do not inflate the garment without wearing it over the intended body area Do not inflate the garment on empty Doing...

Page 7: ...ety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide Equipment not suitable for use in the presence of a flammable anesthetic mixture with air...

Page 8: ...n of voltage to extra low voltage ELV values as provision for basic protection and with no provision for fault protection On console base only in the EU Read instructions before use On garment label L...

Page 9: ...Keep dry On package and console base This side Up On package Keep away from sunlight On package Transportation storage humidity limitation On package Transportation storage atmospheric pressure limita...

Page 10: ...RP Lite Model 760R RP Sports User Manual Do not bleach On garment tag Do not iron On garment tag Label Description Location L98107R0 G 6...

Page 11: ...eases circulation in the areas to which the garment is applied and temporarily relieves muscle aches and pain caused by fatigue or overexertion The RP Lite Massage Cycle The RP Lite console takes air...

Page 12: ...RP Lite Model 760R RP Sports User Manual Figure 1 The Massage Sequence L98107R0 G 8...

Page 13: ...atus indicators Prong Plug see Figure 3 If you use only one garment insert this plug into the unused air outlet on the front of the RP Lite console to seal it and prevent air from escaping DC Power Ad...

Page 14: ...RP Lite Model 760R RP Sports User Manual Caution For indoor use only Use only the DC power adapter provided with the console L98107R0 G 10...

Page 15: ...charged Battery may be charged during normal operation of the console Battery charging time is up to 8 hours Expected service life of battery is two 2 years Replacing the battery is advisable after t...

Page 16: ...Figure 6 Figure 7 Figure 8 and Figure 9 A variety of garments are available Leg garments are available in several sizes RecoveryCore is adjustable Figure 6 Leg Garment Figure 7 Arm Garment Figure 8 M...

Page 17: ...L98107R0 G 13...

Page 18: ...sed to set massage time 15 30 45 60 min or con Time Indicators Indication of set time 15 30 45 60 min or con Start Stop button Used to start or stop massage Battery Status Indicators displays battery...

Page 19: ...ch Used to turn On Off the console located at back of console on the left side DC Adapter Socket Used for connecting DC power adapter located at back of console on the right side Control Function L981...

Page 20: ...RP Lite Model 760R RP Sports User Manual Step 1 Set up the RP Lite Console L98107R0 G 16...

Page 21: ...it onto the pins on the DC power adapter backside see figure below To replace the plug press and hold the Open button under the plug while pulling out the plug see figures below d Insert the DC power...

Page 22: ...P Lite Model 760R RP Sports User Manual Note The console is multi voltage and can be used for travel An appropriate socket adapter must be used in countries with incompatible wall outlets L98107R0 G 1...

Page 23: ...closed at least part of the way up before inserting your leg b Insert your foot into the garment see Figure 11 Pull the top of the garment towards your torso extending the garment to its maximum leng...

Page 24: ...RP Lite Model 760R RP Sports User Manual Figure 11 Putting on the Leg or Mini Boot Garment L98107R0 G 20...

Page 25: ...the RecoveryCore garment Note that the expanders are sided left and right the right side will only fit into the right leg and the left side only into the left leg d Ensure the zippers on the RecoveryC...

Page 26: ...RP Sports Make sure the garment is zipped completely closed before turning on the Power Switch and leave it closed for the duration of the massage At the end of the massage the garment is deflated wi...

Page 27: ...Step 4 Connect the Garment Hose Bundle to the Console L98107R0 G 23...

Page 28: ...arment Hose Bundle to the Console Figure 13 Inserting the Prong Plug to Block the Unused Air Outlet for Single Garment Use Note Hold the console handle with one hand while connecting or disconnecting...

Page 29: ...cycle the console holds the air pressure for a few seconds before releasing the air pausing and then continuing to the next cycle the Hold Time You can set the duration of the Hold Time for your comfo...

Page 30: ...re using two garments You can adjust the time pressure and hold time during massage Using RP Lite for a relaxing and invigorating massage before you exercise increases circulation and can help to get...

Page 31: ...Massage a At the end of the massage turn Off the Power Switch b Wait up to 30 seconds to allow the garment to deflate before you remove it Massage Operation Action Stop Press the Start Stop button Re...

Page 32: ...iate alcoholbased fabric safe disinfectant according to the manufacturer s recommendations When using alcohol based cleaners work in a wellventilated area f Allow the garment to air dry completely bef...

Page 33: ...onsole hose bundle and or garment s is to be disposed of follow the local country regulations taking environmental factors into consideration Caution The console can be transported or stored for short...

Page 34: ...RP Lite Model 760R RP Sports User Manual Replacing the Battery L98107R0 G 30...

Page 35: ...er the carrying handle see figures below b Disconnect the battery and remove it from the battery compartment see figures below c Connect the new battery and then insert it into the battery compartment...

Page 36: ...RP Lite Model 760R RP Sports User Manual 9 Troubleshooting L98107R0 G 32...

Page 37: ...the second one does not The second garment does not receive air Examine its hoses for kinks twists and folds The console stops working the Status Indicator turns On in orange and the alarm beeps Hose...

Page 38: ...for a period of twelve 12 months from date of purchase Garments for a period of six 6 months from date of purchase This warranty does not include or cover malfunctions caused by unreasonable use nonc...

Page 39: ...Distributor in the USA RP Sports 9 Lacrue St Ste 108 Glen Mills PA 19342 855 732 7867 www rpsports com L98107R0 G 35...

Page 40: ...pt at least a distance d 3 3 m away from the system Note The distance calculated from 800 MHz to 2 5 GHz Rated Maximum Output Power of Transmitter W Separation Distance according to Frequency of Trans...

Page 41: ...mestic or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 95 dip in UT for 0 5 cycle 60 dip in UT for 5 cycles 30 dip in UT for...

Page 42: ...rvey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol d 1 2 P d 2 3 800 MHz to 2 5 GHz P d 1 2...

Reviews: