11. EMC Manufacturer Declarations
Electromagnetic Compatibility (EMC) Statement for Home Healthcare
Environment
The RP Lite
®
system has been evaluated to international standard
IEC 60601-1-2 "General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic compatibility -
Requirements and tests".
The Medical Electrical Equipment needs special precautions regarding EMC
and needs to be installed and put into service according to the EMC information
provided here within the accompanying documents.
Portable and mobile RF communication can affect the Medical Electrical
Equipment. See below recommended separation distances between portable
and mobile RF communication equipment and the RP Lite
®
system.
Wireless communications equipment such as wireless home network devices,
mobile phones, cordless telephones and their base stations,
walkie-talkies can affect this equipment and should be kept at least a distance
d = 3.3 m away from the system.
*Note
: The distance calculated from 800 MHz to 2.5 GHz
Rated Maximum Output
Power of Transmitter (W)
Separation Distance* according to
Frequency of Transmitter (m)
0.01
0.23
0.1
0.73
1
2.3
10
7.3
100
23
Caution:
•
Do not use a mobile phone or other devices that emit electromagnetic
fields, near the unit. This may result in incorrect operation of the unit.
•
This system should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, this
machine should be observed to verify normal operation in the
configuration in which it will be used.
Model 760R - Electromagnetic Emissions – Manufacturer’s Declaration
The
Model 760R
is intended for use in the electromagnetic environment specified below. The
customer or the user of the
Model 760R
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The
Model 760R
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
L98107R0-G
36
Summary of Contents for RP Lite Model 760R
Page 2: ......
Page 4: ......
Page 12: ...RP Lite Model 760R RP Sports User Manual Figure 1 The Massage Sequence L98107R0 G 8...
Page 17: ...L98107R0 G 13...
Page 20: ...RP Lite Model 760R RP Sports User Manual Step 1 Set up the RP Lite Console L98107R0 G 16...
Page 27: ...Step 4 Connect the Garment Hose Bundle to the Console L98107R0 G 23...
Page 34: ...RP Lite Model 760R RP Sports User Manual Replacing the Battery L98107R0 G 30...
Page 36: ...RP Lite Model 760R RP Sports User Manual 9 Troubleshooting L98107R0 G 32...