background image

2.3. Maintenance

This product is a precision RF Shielding Enclosure manufactured to be sturdy and robust. In order to ensure the optimal

performance of the product, periodic inspection is required according to the number of uses and period of use.

2.3.1 Recommended Routine Maintenance and Method

* Check to ensure that the Module, Hinge, and Latch screws are not loose and tighten if necessary.

* Check to ensure that the torque of the hinge mounted on the door has not weakened. The hinge has passed the

30,000 open-close private cycle test, but the torque may weaken as the number of openings and closings increases.

However, weakening of hinge torque does not affect the shielding effectiveness.

If the torque is weak, the lid could shut abruptly at a certain point, so please be careful as it may be hazardous. If

necessary, please contact customer support or an agent for replacement. It can be replaced by the customer, but

special attention and consultation are required as it may affect the shielding effectiveness.

* Check for damaged cable harnesses and replace them if necessary.

* Shielding performance may be greatly reduced if metal corrosion or oxidation progresses on the surface that the

gasket touches on the door or on the gasket frame. Clean thoroughly with alcohol and a soft cloth. It must be kept

clean at all times.

* Typically, the shielding effectiveness performance of new products has passed

10,000

open-closed private cycle

tests. However, it is recommended to replace the gasket if the shielding performance does not improve even after

regular maintenance has been performed.

* To maintain the shielding effect, refer to the performance test on page 7

2.2. Performance Test

. It is recommended

to perform a test approximately every 5,000 times or at intervals of 6 months.

2.3.2 Recommended EMI Gasket replacement Period

* Generally, after opening and closing the door about 30,000 times, the gasket's contact surface may be damaged. In

this case, perform routine maintenance first, and if it does not improve, replace the gasket.

* Generally, long-term storage of the product with the door closed may affect the elasticity of the foam gasket to

decrease even if it is used less than the recommended number of lid openings and closings.

In this case, perform routine maintenance first and keep the door open for more than one day, and then perform

the performance test. If there is no improvement, replace the gasket.

RedwoodComm Co., Ltd.  |

RWC7100A Operation Manual

Page 10 of 16

Summary of Contents for RWC7100A

Page 1: ...RWC7100A RF SHIELDING ENCLOSURE Operation Manual R211216_1 10...

Page 2: ...mance Test 7 2 2 1 How to Set up the Equipment Required for RF Shielding Effectiveness Test 7 2 2 2 Test Procedure 8 2 2 3 An Example Test Result of RF Shielding Effectiveness 9 2 3 Maintenance 10 2 3...

Page 3: ...nished or recommended by RedwoodComm causes problems or damage that is attached to or used in connection with the product Where any additional parts equipment that is not authorized by RedwoodComm was...

Page 4: ...customization panel To maintain lasting shielding effectiveness users can order additional EMI gaskets and easily replace them when necessary There are many kinds of wireless standards in various ind...

Page 5: ...effectiveness please refer to the 2 3 Maintenance Dimensions and Weight Dimensions W x D x H mm inches Interior workspace 260 x 190 x 100 10 2 x 7 5 x 3 9 Exterior 355 x 270 x 185 14 0 x 10 6 x 7 3 We...

Page 6: ...tdoor and where the ground is not flat If this product is not used for a long time store it in a dry place away from direct sunlight covered with vinyl or placed in a cardboard box If this product is...

Page 7: ...oor from the top 4 x M3 Hole DP 8mm Mounting position for RWC8201A only EMI Mesh Gasket This detachable part is for obtaining optimal RF shielding 2set x EMI Foam Gasket This detachable part is for ob...

Page 8: ...capable of generating 400M 6400MHz tone signal Signal generator side antenna omni directional one Signal power meter A spectrum analyzer or equivalent device that can measure the tone power of a signa...

Page 9: ...gnal generator or RWC1020A as follows Frequency 400 1MHz Power 10dBm 5 Open the RWC7100A door and measure the power Pref moving the reference antenna to several points as shown in Figure 2 3 Figure 2...

Page 10: ...RF Shielding Effectiveness The product conformity test was measured using a method modified from the IEEE STD 299 1 standard Figure 2 5 Shielding effectiveness of RWC7100A RedwoodComm Co Ltd RWC7100A...

Page 11: ...rosion or oxidation progresses on the surface that the gasket touches on the door or on the gasket frame Clean thoroughly with alcohol and a soft cloth It must be kept clean at all times Typically the...

Page 12: ...of the 1st Mesh Gasket place the seam in the center and insert it slowly with an even density 5 In the case of the 2nd and 3rd Foam Gasket refer to the picture above remove the tape on the back of the...

Page 13: ...Set RWC8201A mounted on the lid of RWC7100A Antenna Dimensions W x D x H mm 130 x 130 x 15 Frequency Range 700MHz to 8 5GHz N F to SMA F Adaptor and RF cables installed O7100A 92 O7100A 93 Figure 3 1...

Page 14: ...use shielded cables to ensure shielding effectiveness 5630 2000 D Sub 9Pin Module Type B Outside dimensions W x D x H mm 102 x 80 x 24 DB 9P M outside to DB 9S F inside 5 000pF Pi Filter Unused ports...

Page 15: ...rformance Test after installation 3 2 Example of ordering This is an example of how customers order the RWC7100A with IO Modules For more information please contact our sales or distributor before pla...

Page 16: ...the RWC8201A antenna is measured not only in the 400 MHz band but also in the 900 MHz band Especially transfer function between antennas in the 400 510 MHz and 860 960 MHz bands are dealt with in det...

Page 17: ...RWC8201A antennas is about 40dB and starting decreasing from 490MHz it becomes 27dB at 510MHz In the 900MHz band which is within the recommended frequency range of RWC8201A antenna the path loss is ab...

Page 18: ...A 2 DB9 Module Pin map Fig A 4 DB9 Module Pin Map RedwoodComm Co Ltd RWC7100A Operation Manual Page 17 of 16...

Reviews: