9
Changing the Light Meter unit of Measure
1. Press and hold the
UnIT
button, while in the Light Meter
measurement mode.
2. The LCD display will toggle between the units LUX and Ft-cd.
3. Release the UnIT button to select the displayed unit.
4. The selected unit will be saved as the new default.
Zero Adjustment
1. While taking a light measurement, completely cover the light sensor.
2. If the LCD display does not show value of zero, press the and
buttons simultaneously until the LCD display shows a zero value.
Type K/J Thermocouple Measurement Mode
1. Select the "tP" function (see
Selecting Measurement Modes
for details).
2. Plug the optional Type K or J Thermocouple Probe into the Type K/J
thermocouple input jack.
3. The screen will display a "K", confirming that is configured by default
to a Type K thermocouple probe. The measured temperature will now
be displayed on the display screen. As shown below.
Note:
If using a type J thermocouple probe, see
Selecting Measurement
Modes
for details.
K
continued...