reedinstruments
www
com
5
Measurement rate:
2 times per second, nominal
Auto power off:
Meter automatically shuts down after
approx. 30 minutes of inactivity
Operating environment: 0 to 50°C (32 to 122°F)
at < 70% relative humidity
Storage temperature: -20 to 60°C (-4 to 140°F)
at < 80% relative humidity
Power Supply:
Standard 9V battery (NEDA 1604, IEC 6F22)
Dimensions:
197 x 88 x 41mm
Weight:
635g (with holster)
Accuracy
Accuracy is given at 18°C to 28°C (65°F to 83°F), less than 70 % RH
DC Voltage (Auto-ranging)
Range
Resolution
accuracy
400.0mV
0.1mV
±0.5% of rdg + 2 digits
4.000V
1mV
±1.2% of rdg + 2 digits
40.00V
10mV
400.0V
100mV
1000V
1V
±0.8% of rdg + 2 digits
Input Impedance: 50 MΩ & 400 mV range; 8.2 MΩ & other ranges
Maximum Input: 1000 V DC or 1000 V AC RMS.
aC Voltage (Auto-ranging except 400 mV)
Range
Resolution
accuracy
400.0mV
0.1mV
±2.0% of rdg + 3 digits
4.000V
1mV
±0.8% of rdg + 3 digits
40.00V
10mV
400.0V
100mV
750V
1V
±2.0% of rdg + 4 digits
Input Impedance: 50 MΩ & 400 mV range; 8.2 MΩ & other ranges
Frequency Range: 40 to 400 Hz