Version 10/21
Page 29 of 49
PAWS UM EN
Figure 21. Opening and adjusting the width of the Clamp Jaws.
•
Adjust the lateral angle of the Clamp Jaws to the front tubes. Adjust the Clamp Jaw angle to the
Wheelchair tube by rotating at Bolt A. The Clamp Jaws should grip the Wheelchair tube evenly when
they are closed. Do this for both sides.
Attention!
The recommended Torque setting of Bolt A is 50 Nm (Newton Meters).
Figure 22. Changing the Lateral Angle and Depth of the Clamp Jaws
•
Setting the Frontal Angle of the Clamp Jaws to the front tubes. Rotate the Clamp Jaws so that they are
parallel to the tubes.
Figure 23. Changing the Frontal angle of the Clamp Jaws
•
Setting the depth of the Clamp Jaws to the Device. Adjust the Clamp depth (determines distance from
Device to Wheelchair) to ensure space between the User
’
s knees and the Device.
Figure 24. Depth adjustment of the Clamp mechanism in relation to the Device
A
B