D11
Power supply connections:
Up to three lamps may be supplied from the power
supply shown, with connectors for the following
lamps:
⚫ 100W low-voltage halogen lamp
⚫ HBO 200W/4 or CS 200W/4 mercury-vapor high-
pressure lamp
⚫ CSI 250W/1 metal halide short-arc lamps
⚫ Transformer for 220v line voltage, remote-control
connector for the XBO 450W xenon lamp.
⚫ The housing has a socket ➀ for the line power
connection on the back.
Supply for the 100W low-voltage halogen lamp
2. Volt meter
3. Indicator
4. Rotary dimmer control for the low voltage lamp
and zoom lighting
5. Connector for zoom lighting
6. Connector for 6V low voltage lamp
7. Connector for 12V low voltage lamp
⚫ The control transformer serves the 6V or 12V low-
voltage lamps. It must be connected only to
alternating current. The primary voltage range is set
by 100-125V or 200-250V in the factory. The line
cord into the power supply supplies power to
everything.
⚫ When you turn on the rotary switch
➃
, power is
supplied to the microscope base to light up the
lamp; more CW, the greater the lamp brightness.
⚫ The upper scale of the voltmeter is for 12V, the
bottom for 6V. The pointer of the voltmeter must not
exceed the red marks at 12V or 6V.
⚫ The transformer from the socket box is removed
to replace the miniature fuse. This can be done after
unscrewing the four fixing screws
➇
on the back of
the cabinet. For 200-250V primary voltage, the fuse
should be 0.8A; at 100-125V, the fuse should be
1.25A.
If a different control transformer is used in the power
supply, the back will have only the 12V 100W low-
voltage halogen lamp connection