D14
Power supply for 450W/2 xenon lamp
1. test button for voltage
2. voltmeter
3. current controller
4. ammeter
5. main switch
6. fuse, 8A
7. pilot light
8. fuse, 1A
9. connection for control cable for the remote
control unit or jumper
10. connection for connection cables to the lamp
house 50
11. socket for power cord
12. hour counter
⚫ The control unit can be operated only with
220-240V line voltage. The appropriate intermediate
transformer must be used for different voltages.
⚫ The cable from the lamphouse 50 with special
plug is connected to the socket
➉
.
⚫ The power cord is connected to the socket 11.
⚫ A jumper must be inserted in the outlet
➈
if no
remote control unit is connected.
⚫ Turning on the lamp is done with the main switch
➄
, whereby the pilot lamp
➆
lights up and the fan in
the lamphouse 50 comes on. Then, the lamp is lit
with the red start button on the lamphouse 50. See
instructions on page D10!
⚫ The current controller is set at the factory for 25A
and should not be adjusted.
Remote control device for the control unit to
the XBO 450W/2 Xenon lamp
13. level regulator for electric current (17-30A)
14. hour counter
15. power switch
16. Control cable socket
⚫ The remote control unit is mounted in the rack
house described on page D11.
⚫ The connection between the power supply for the
xenon lamp and the remote control device is via the
control cable. This is connected to the power supply
instead of the short circuit plug to the socket
➈
, and
on the remote control unit to the line socket 16.
⚫ Lighting up the lamp is done by the power switch
15; which position the main switch
➄
of the control
unit is in, does not matter. Please follow the notes on
page D10 when the lamp is switched on!
⚫ The brightness of the lamp can be changed with
the control knob 13