30
6
To mark the test as a control test, Press and
Release the
button once.
shows on
the display window.
Important:
If you do not mark the test as a
control test, it will be saved in your monitor’s memory
as one of your personal blood glucose results. This could
affect your blood glucose averages.
Applying Control Solution to the Test Strip
1
Turn the control solution bottle upside down
three to four times to mix the solution.
2
Remove the cap.
3
Apply a drop of control solution
to the test strip in the area
shown here. The control solution
is drawn into the test strip.
4
Continue to touch the control solution to the test
strip until the monitor begins the test.
The monitor begins the test when:
•
You hear the beeper, if the beeper is ON.
•
The display window shows the status bar.
•
Then the display window shows the countdown.
Status bar
Countdown
Contr
ol
Solution
Testing
ART21258_Rev-A-Man.indd 33