31
Note: Do not
remove the test strip from the monitor
or disturb it during the countdown.
Important:
If the countdown does not start:
What It Means:
You might not have applied enough
control solution to the test strip.
What to Do:
Apply a second drop of control solution
to the test strip. Refer to your test strip
instructions for use for the number of
seconds you have to apply a second drop.
•
If the countdown still does not start, or if the number
of seconds you have to apply a second drop have
passed, discard the test strip, turn off your monitor,
and try again with a new test strip.
5
At the end of the countdown:
•
If the beeper is ON, listen for the beeper.
•
The control result shows on the display window.
•
The result is stored in your monitor’s memory
as a control result. In your logbook, record the
result as a control result.
Shutting Off Your Monitor
1
Removing the test strip from the
strip port turns off the monitor.
You can use the opened foil
packet to remove and discard
your used test strip.
Contr
ol
Solution
Testing