39
PROBLEM
POSSIBLE
CAUSE
WHAT YOU SHOULD DO
The outer ENERGY
LEVEL lights are
flashing alternately
and the i-Light®
Pro system is
beeping every ½
second.
The i-Light®
Pro
system isn’t
working
properly.
Contact Customer Service (see page 44)
for help.
The second and
fourth ENERGY
LEVEL lights are
flashing and the
system is beeping.
The SKIN
CONTACT
SENSORS may
be stuck.
Lift the HAND PIECE from the BASE UNIT and
check the SKIN CONTACT SENSORS. The
CONTACT SENSORS should slightly protrude
from the sides of the HANDPIECE. You should
be able to press the CONTACT SENSORS down
so that they are flush with the HAND PIECE.
They should spring back to about 2mm when
released. If the CONTACT SENSORS do not
depress or spring back they may be stuck and
may require cleaning. See page 32 for instruc-
tions on how to clean your i-Light® Pro IPL
system. If you still can’t get it clean, contact
Customer Service on page 44 for help.
Troubleshooting