To maintain the air quality:
1. Ensure the air provided to the filtration system meets the specification given.
Incoming air supply specification:
Air pressure
5 bar - 8.5 bar*
Air consumption
Up to 50 litres per minute
BS ISO 8537-1:2010 [4:4:4]
Particle size
15 μm
Dirt concentration
8 mg/m
3
Dew point
+3 °C
Oil
5 mg/m
3
* NOTE:
The exact values are dependent on the inherent characteristics of the air pipe between the air filters and the REVO-2 head.
Please ask the installer for the specification.
If the air pressure falls outside the range given, the head metrology results will be affected.
2. Regularly monitor the dew point indicator (see image below) and follow the instructions given in the table.
For more information please visit the air filtration manufacturer's website
and search for IDG5 filters. (Please
refer to the section 'REVO air filter service kits' for details on air filter service kit part numbers).