REVO-2 probe specification
RSP2 and RSP2 V2
RSP3-x
SFP2
Used for
High-speed 5-axis 2D scanning *
and single point taking
3-axis traditional CMM scanning
and single point taking
Surface finish measurement
Stylus range
M2 but range has been selected
to opimise high speed scanning
performance **
M3
SFS-#
SFM-#
Cranked styli
No
Yes
Yes
Crash protection
XY and –Z via break off of either
module or stylus holder, +Z via
integral bump-stop design
XY and –Z via break off of either
module or stylus holder, +Z via
integral bump-stop design
XY and –Z via break off of holder
Calibration
Uses Ø30 mm sphere supplied
with REVO‐2 or optional Ø45
mm sphere
Uses Ø30 mm sphere supplied
with REVO‐2 or optional Ø45
mm sphere
Uses SFA# and TFP
* All scans should be configured to allow an angle no greater than 45° between the measured surface and the axis of the RSP2 or RSP2 V2
probe.
** For more detail refer to UCCserver 5.0 (or newer) or PBC-01909.
NOTES:
For optimum metrology, it is essential that all probe and stylus holder changes are carried out automatically through the
appropriate change ports.
Avoid touching the joint faces, as they will become contaminated. It is vital that the joint surfaces are protected and stay clean.
Replace covers when not in use.