Calibration
System calibration
The REVO-2 datum sphere is used to calibrate the REVO-2 head as well as RSP2 and RSP3-x probe configurations.
The SFP1 is calibrated using the SFA that is fixed to the MRS rail.
For the minimum required CMM volume please refer to the table and diagram below:
Probe type
A ﴾in mm﴿ ‐ Ø30 sphere
A ﴾in mm﴿ ‐ Ø45 sphere
RSP2 with RSH175
420*
435*
RSP2 with RSH250
570*
585*
RSP2 with RSH350
770*
785*
RSP2 with RSH450
970*
985*
RSP2 with RSH500
1070*
1085*
* For a Ø6 mm × 10 mm stylus.
Dimensions in mm
REVO-2 user's guide
www.renishaw.com
Issued 07 2019