d) Tilt the upper end of the UV lamp chamber out.
e) Wear gloves included in the new UV lamp package. Avoid direct skin contact with the quartz
glass of the UV lamp. Remove the UV lamp cover to expose the UV lamp. Unplug the UV lamp
from its power cord, and then carefully remove the old UV lamp from the chamber.
f)
Carefully insert the new UV lamp into its chamber. When about 2/3 of the lamp is inserted, hold
the UV lamp and connect it to the ballast cable connector (4-pin connector) as shown in the
picture, and then gently insert the new UV lamp completely into the chamber.
g) Cap the UV chamber with the black mask.