KrosFlo® TFF System
–
Auxiliary Scale ACSS-20K
User Guide
24 repligen.com 420-14871-001 Rev. 00
13.1.1
Working mode selection
Parameter allowing to select method of determination of sample piece mass.
1.
Enter
<3.2.PcS / 3.2.2.UUt>
submenu.
2.
Go to
<Acc>
function.
3.
Press
key, parameter values are displayed successively one by one, where:
S_S
–
Select
to set sample mass by determining mass of a single part,
Suu
–
Select to set sample mass by
entering mass of a single part.
4.
Enter respective value and press
key to confirm, then continue weighing.
13.2
Setting reference sample mass by entering mass of a single part
1.
Enter
<3.2.PcS / 3.2.2.UUt>
submenu, set
<Suu>
value.
2.
Enter
<PcS>
working mode (Parts Counting), first, text
<Set_Ut>
is displayed for 1s, next you
see a window for entering mass value of a single part.
3.
Enter respective value and press
key to confirm, the home screen is displayed
automatically along with the quantity of parts loaded onto the weighing pan (pcs).
Note:
If the value of entered single part mass is greater than max capacity value, then message <Err
Hi> is displayed.
13.3
Setting reference sample mass by determining mass of a single part
1.
Enter
<3.2.PcS / 3.2.2.UUt>
submenu, set
<S_S>
value.
2.
Enter
<PcS>
working mode (Parts Counting), first, text
<Set_Ut>
is displayed for 1s, next you
see a window for entering mass value of a single part.
3.
Press
key to select one of the following options:
10
–
10 pieces,
20
–
20 pieces,
50
–
50
pieces,
100
–
100 pieces,
0000
–
Custom reference sample quantity to enter required value.
4.
Select respective option and press
key to confirm, first, text
<LoAd>
is displayed for 1
second, next you see a weighing window.
5.
If the parts are to be weighed in a container, first put the container on a weighing pan and
tare it.
6.
Load the weighing pan with the declared amount of parts. When the indication is stable (
pictogram is displayed), press
key to confirm the mass.
7.
Single part mass is calculated automatically, the quantity of parts (pcs) is displayed.
Note:
Total weight value of all parts loaded onto the weighing pan cannot be greater than the max
capacity value.
Note:
Single part mass value must be equal or greater than 0.1 of the reading unit. Unless this
condition is met, the weighing device displays a message
<ErrLo>
.
Note:
In the course of parts quantity determination it is necessary to wait for a stable measurement
before confirming the declared quantity value.