KrosFlo® TFF System
–
Auxiliary Scale ACSS-20K
User Guide
18 repligen.com 420-14871-001 Rev. 00
7.1.5
Port
1.
Enter
<P5.2.Prtr / 5.2.1.Prt>
submenu.
2.
Press
key, options are displayed successively, where:
nonE
–
No port selected,
rS1
–
RS232 (1),
USbA
–
USB type A which a printer can be connected, and
USbb
–
USB type B
which a computer can be connected.
3.
Set respective value and press
key to confirm, next go to the home screen.
7.2
Additional display
<5.3.AdSP>
submenu allows to:
•
Select port to which the additional display is connected
7.2.1
Port
1.
Enter
<P5.3.AdSP / 5.3.1.Prt>
submenu.
2.
Press
key, options are displayed successively, where:
nonE
–
No port selected,
rS1
–
RS232 (1).
3.
Set respective value and press
key to confirm, next go to the home screen.
8.
Printouts
8.1
Calibration report
<P6.1.CrEP>
is a group of parameters allowing to declare variables that are to be printed on the
calibration report printout. Each variable allows for accessibility attribute:
YES
–
print,
no
–
do not
print. Calibration report is automatically generated at the end of each calibration process.
Table 6. Calibration report variable list
Submenu
Name Description
6.1.1.
CtP
Calibration type
6.1.2.
dAt
Calibration date
6.1.3.
tin
Calibration time
6.1.4.
Idb
Scale serial number
6.1.5.
CdF
Difference between mass of a calibration weight measured during the last
calibration and mass of the currently measured calibration weight
6.1.6.
dSh
Dashed line separating printout data and signature fields
6.1.7.
SiG
An area for the signature of an operator carrying out the calibration