KrosFlo® TFF System
–
Auxiliary Scale ACSS-20K
User Guide
25 repligen.com 420-14871-001 Rev. 00
14.
Working mode
–
+/- control
+/- control is working mode enabling to enter check weighing threshold values (
Min
,
Max
).
14.1
Local settings
To go to local settings, enter
<3.3.HiLo>
submenu.
Table 14. +/- control
–
Local settings
Submenu
Setting
3.3.1.Acc
Working mode accessibility
3.3.2.Snn
Save mode
3.3.3.Int
Time interval
3.3.4.Lo
Lo threshold
14.2
Declaring check weighing thresholds
1.
Enter
<HiLo>
working mode (+/- control), first, text
<SEt Lo>
is displayed for 1 s, next you see
a window for declaring low weighing threshold (Min).
2.
Enter a respective value and press
key to confirm, first, text
<SEt Hi>
is displayed for 1
second, next you see a window for declaring high weighing threshold (Max).
3.
Enter respective value and press
key for confirmation, working mode’s home screen is
displayed along with the declared threshold value, where:
Min
–
Load mass lower than low
weighing thresholds,
Ok
–
Load mass within weighing thresholds,
Max
–
Load mass greater
than high weighing threshold.
Note:
If the entered low threshold value (Min) is greater than the high threshold value (Max), <Err
Lo> error is displayed
.
Note:
If the entered high threshold value (Max) is greater than the maximum capacity value,
<Err Hi>
error is displayed.
15.
Working mode
–
Percent Weighing
Percent Weighing is a working mode enabling to compare measured load mass with the reference
sample mass. The result is expressed in [%]. Reference sample mass can be either determined by
weighing or entered to weighing device memory by an operator.
15.1
Local settings
To go to local settings, enter
<3.4.dEu>
submenu.