KrosFlo® TFF System – Auxiliary Scale ACSS-60K
User Guide
24 repligen.com 420-14871-002 Rev. 00
13.
Working mode – Parts Counting
Parts Counting is a working mode enabling to determine quantity of small pieces of the same mass,
which determination is done on the basis of mass of sample piece (single part), and where the
sample piece mass (single part mass) is determined using the weighing device.
13.1
Local settings
To go to local settings, enter
<3.2.PcS>
submenu.
Table 13. Parts Counting – Local settings
Submenu
Setting
3.2.1.Acc
Working mode accessibility
3.2.2.UUt
Operation mode
3.2.3.Snn
Save mode
3.2.4.Int
Time interval
3.2.5.Lo
Lo threshold
13.1.1
Working mode selection
Parameter allowing to select method of determination of sample piece mass.
1.
Enter
<3.2.PcS / 3.2.2.UUt>
submenu.
2.
Go to
<Acc>
function.
3.
Press
key, parameter values are displayed successively one by one, where:
S_S
– Select
to set sample mass by determining mass of a single part,
Suu
– Select to set sample mass by
entering mass of a single part.
4.
Enter respective value and press
key to confirm, then continue weighing.
13.2
Setting reference sample mass by entering mass of a single part
1.
Enter
<3.2.PcS / 3.2.2.UUt>
submenu, set
<Suu>
value.
2.
Enter
<PcS>
working mode (Parts Counting), first, text
<Set_Ut>
is displayed for 1s, next you
see a window for entering mass value of a single part.
3.
Enter respective value and press
key to confirm, the home screen is displayed
automatically along with the quantity of parts loaded onto the weighing pan (pcs).
Note:
If the value of entered single part mass is greater than max capacity value, then message <Err
Hi> is displayed.
13.3
Setting reference sample mass by determining mass of a single part
1.
Enter
<3.2.PcS / 3.2.2.UUt>
submenu, set
<S_S>
value.
2.
Enter
<PcS>
working mode (Parts Counting), first, text
<Set_Ut>
is displayed for 1s, next you
see a window for entering mass value of a single part.
3.
Press
key to select one of the following options:
10
– 10 pieces,
20
– 20 pieces,
50
– 50
pieces,
100
– 100 pieces,
0000
– Custom reference sample quantity to enter required value.