KrosFlo® TFF System – Auxiliary Scale ACSS-60K
User Guide
27 repligen.com 420-14871-002 Rev. 00
4.
Enter respective value and press
key to confirm. The home screen is displayed
automatically with
0.000%
value.
Note:
If the value of entered reference sample mass is greater than the max capacity value, then
message
<Err Hi>
is displayed.
16.
Working mode – Peak Hold
Peak Hold is a working mode allowing to snap value of maximum force applied to the weighing pan
during one weighing process.
16.1
Local settings
To go to local settings, enter
<3.5.toP>
submenu.
Table 16. Peak Hold – Local settings
Submenu
Setting
3.5.1.Acc
Working mode accessibility
3.5.2.Lo
Lo threshold
16.2
Peak Hold operation
1.
Enter
<3.5.toP / 3.5.2.Lo>
submenu, set
<Lo>
parameter value (Lo threshold) after
exceeding of which maximum force is to be registered.
2.
Enter
<toP>
working mode (Peak Hold). From now on the scale registers and holds every
single weighing which is above the
Lo threshold
, and which is higher than the result of the
previous Peak Hold. Snapped Peak Hold value is signaled by
<Max>
pictogram at the top of
the screen.
3.
The start of the next process of Peak Hold measurement is possible only after removing the
load form the weighing pan and pressing
key.
4.
This causes returning to the home screen of
<toP>
mode. Pictogram
<Max>
is automatically
deleted.
17.
Totalizing
Totalizing is a working mode enabling to sum mass of all weighed ingredients, and to print (via scale-
connected printer) the total mass value. The pictogram allows to sum up to 30 weighings
(ingredients) maximum within one process.
17.1
Local settings
To go to local settings, enter
<3.6.Add>
submenu.