Accupak Rechargeable
Battery System
(Optional Accessory)
8
9
Please read these instructions if you have an Accupak rechargeable battery system
for mains free usage.
Components received on delivery / Set up
• The Accupak battery
will be fitted out of sight on the base of the chair.
• The transformer and power cables
which enable the battery to be initially charged /
recharged will normally be connected to the chair when it is initially set up.
• The handset user instructions
which can be found on page 15, 16 and 17.
• Only use the equipment supplied with the Accupak
.
Operation and charging
•
When the chair is received the battery
MUST BE FULLY CHARGED
before usage. It is
recommended that the battery is
CHARGED FOR AT LEAST 8 HOURS
.
•
The battery
SHOULD BE OPERATED
at normal room temperatures and never in damp
environments.
•
The battery provides an average of up to 70 adjustment cycles.
•
The plug-in
CONNECTIONS ARE ACCESSIBLE
so that the system can be switched off in
an emergency.
•
The transformer must
ALWAYS
lie flat on the floor.
•
The Accupak has an audible acoustic warning signal which is activated when the handset controls
are used and the battery is low and requires recharging. The system still remains activated and
emergency adjustment is possible, but please charge immediately. The battery also has an overall
switch off threshold should you continue to operate the chair when the acoustic warning is
sounding. This prevents deep discharge of the battery and permanent damage to it.
• DO NOT
let the battery completely discharge as this will stop it from being fully charged again.
If the battery is left to completely discharge it may have to be replaced and this is
not
covered under the warranty
.
•
It is
RECOMMENDED
that the battery is charged regularly overnight to prevent complete
discharge.
•
The chair may be used and operated whilst it is being charged, but
DO NOT
move the chair
about when it is plugged in.
•
The electrical components are designed to be used for
NO
LONGER
than 2 minutes continuous
movement in any period of 20 minutes. (See Usage time on page 17).
•
Rechargeable batteries, by their nature, have limited lifespans and are characterised by gradual
power loss over time. This normal depreciation and continual power loss is
not covered under
the warranty
.
Accupak Rechargeable
Battery System
(Optional Accessory)
Problem
Possible Cause
Remedy
Handset or mechanism
movement not working
(mains power
disconnected from chair)
1) No supply voltage
2) Deep discharge protection
triggered (see above)
3) Handset or system defective
• Connect chair to the mains/
charge the battery
• If connecting to the mains
/charging the battery does
not work, contact supplier/
engineer
Handset or mechanism
movement not working
(mains power connected
to chair)
1) Thermoswitch has been
triggered in transformer (you
cannot see this), preventing the
chair from being powered and
the battery from charging
2) Handset or system defective
• Leave the chair in a state of
rest for 20 to 30 minutes
• If leaving the chair at rest
does not work, contact
supplier / engineer
Acoustic Warning Sound
can be heard
1) The battery charge is very low
(see above)
• Immediately recharge the
battery
Maintenance and disposal
•
The rechargeable battery system is designed to be maintenance free.
• DO NOT
open the Accupak. Check the cables and connections regularly for any signs of damage.
• DO NOT
operate the system if there is any damage or if the unit gets too hot.
• MAKE SURE
that the Accupak is kept clean and dry at all times.
• DO NOT
submerge in water or expose to direct heat.
• NEVER
run anything over the cables e.g. a vacuum cleaner, as this may damage them.
•
ALWAYS
store the chair in a dry and safe place which will protect the chair and the Accupak
battery. If you are storing for periods of more than 6 months then the Accupak battery must be
fully charged before being stored. The battery should be re-charged every 12 months when in
storage.
•
If your battery shows any signs of leakage
DO NOT
allow the fluid to come into contact with
your skin. If it does, clean the affected area and contact your doctor.
•
ONLY
clean the Accupak and transformer by wiping carefully with a slightly damp cloth.
Do not use solvents or chemical cleaning agents, to clean them.
ALWAYS DISCONNECT
the power source when doing any cleaning
.
•
Damaged batteries must be disposed of immediately and in line with environmental regulations.
Basic fault and error detection
Below we have listed some basic checks you can carry out if the chair mechanism stops working. If
the chair is still not working after carrying out these steps we recommend that you contact
your supplier.