Alarms
English 109
Setting the alarm volume
The volume level of the Astral device can be set from one to five (with five being the loudest and the
default being three).
It is also possible to set a minimum volume level, where any settings below the minimum are greyed out
and disabled from use. This feature is configurable in the Clinical menu only.
WARNING
•
Alarm volume cannot be adjusted separately for individual alarms. Adjusting the alarm volume
will change the volume of all alarms, independent of alarm priority level.
•
When adjusting alarm volume, ensure that the alarm can be heard above the ambient noise
levels that the patient may experience in a variety of settings, including use in noisy
environments or inside mobility bags.
In the example below the current alarm volume is '5' however the minimum alarm volume has been set
at '3'. The '1' and '2' volume options are now disabled and not able to be selected by the patient or carer.
Testing the alarm sounders and indicators
To confirm the alarm will sound as intended, regularly test the alarm.
The Astral device incorporates two alarm sounders. During an alarm condition both sounders are
operated in parallel. To confirm the correct operation of each sounder, regularly perform the Alarm test
function. During this test each sounder will be operated separately and in sequence.
To test the alarm sounders and indicators:
1.
Press
. The Alarms screen is displayed.
2.
Press
. The Alarm volume screen is displayed.