Data management process
128
Data management summary
The table below summarises the data available for use in ResScan.
Detailed data
Logged for the last 7 days
Summary data
1
(5
th
, Median, 95
th
)
Logged for the last 365 days
Pressure
25 Hz
Flow
25 Hz
Volume
25 Hz
PIP
1 Hz
PEEP
1 Hz
Avg. P
1 Hz
Maximum Inspiratory flow
1 Hz
Leak
1 Hz
SpO
2
1 Hz
Pulse
1 Hz
FiO
2
1 Hz
Vt
1 Hz
MV
1 Hz
Va
1 Hz
Resp. rate
1 Hz
I:E ratio
1 Hz
Ti
1 Hz
Te
1 Hz
% Spont. trig
% Spont. cyc
RSBI
1 Hz
AHI
2
Median
AI
2
Median
1
One set of summary data is provided per day for each program used.
2
The Apnoea Index (AI) and Apnoea Hypopnoea Index (AHI) are only recorded when a single limb circuit with intentional leak is
used. AI indicates the number of times per hour that patient flow reduces by more than 75% of nominal flow for a period of
10 seconds or more. AHI indicates the number of times per hour that patient flow reduces by more than 50% of nominal flow for
a period of 10 seconds or more.