Cleaning and maintenance
English 137
Device information
Device information, including the Next service due date indicating when the next preventative
maintenance is due, can be found by pressing
and selecting Device.
Upgrading software
Software upgrades can only be performed by an authorised service representative. Astral has the ability
to be upgraded with a USB stick containing the appropriate software. Please contact your authorised
agent for further information.
Additional considerations for hospital or healthcare facilities
Component
Hospital or healthcare facility
Device
Follow the cleaning and maintenance protocol and schedule of the hospital/healthcare facility.
Double limb adapters
For general hygiene, the double limb adapter should be replaced between patients or protected with an
antibacterial filter.
Air filter
Replace every six months (or more if necessary).
Mask
Masks should be reprocessed or replaced between patients. Cleaning, disinfection and sterilisation
instructions are available from the ResMed website, www.resmed.com/downloads/masks. If you do
not have internet access, please contact your ResMed representative.
Patient circuits
Follow the protocol of the hospital/healthcare facility and the recommendations of the manufacturer.
Humidifier
Follow the protocol of the hospital/healthcare facility and the recommendations of the manufacturer.
Antibacterial filter
Replace as required and between patients.