Technical specifications
English 139
Dimensions (L x W x H)
285 mm x 215 mm x 93 mm
Display screen size: 150 mm x 90 mm
Weight
3.2 kg
Inspiratory port / double limb adapter
22 mm taper, compatible with ISO 5356-1:2004
Anaesthetic & Respiratory Equipment – Conical Connectors
Pressure measurement
Internally mounted pressure transducers
Flow measurement
Internally mounted flow transducers
Power supply
AC 100–240V, 50–60Hz, 90 W 3.75 A continuous, 120 W / 5A peak
110V/400 Hz
External DC Power Supply
12 - 24V DC 90 W, 7.5 A / 3.75 A
Internal Battery
Lithium-Ion battery, 14.4 V, 6.6 Ah, 95 Wh
Operating hours (standard case): 8 hours with a new battery under normal
conditions
Test conditions: Adult, (A)CV mode, Vt = 800 mL, PEEP = 5 cmH
2
O,
Rate = 20 bpm, Ti = 1.0 sec. All other parameters remain at default
settings.
Operating hours (worst case) > 4 hour run time under the following
conditions:
Test conditions: Adult, non-vented, PACV mode, Double limb circuit,
Pressure Assist = 30 cmH
2
O, PEEP = 20 cmH
2
O Rate: 20 bpm,
Ti: 1.0 sec, Rise Time = Min, Safety Vt = Off, Trig = Off. All other
parameters remain at default settings.
Total lifetime: up to 3,000 hours of normal operation on internal battery
Note: Time may vary with different settings and environmental conditions.
Housing construction
Flame retardant engineering thermoplastic
Environmental conditions
Operating temperature: 0°C to 40°C
Charging temperature: 5°C to 35°C
Operating humidity: 5 to 93% non-condensing
Storage and transport temperature: -25°C to 70°C for up to 24 hours
Storage and transport temperature: -20°C to 50°C for greater than 24 hours
Note:
Storing the Astral device at temperatures exceeding 50°C for
extended period of time may accelerate battery aging. This will not affect
the safety of the battery or device. Refer to Using the internal battery (see
page 63)
Storage and transport humidity: 5 to 93% non-condensing
It takes 40 minutes* for the device to be ready for use on a patient when
removed from storage at the minimum long term temperature and at an
ambient temperature of 20°C.
*Assumes that the device is connected to an external AC power.
It takes 60 minutes for the device to be ready for use on a patient when
removed from storage at the maximum long term temperature and at an
ambient temperature of 20°C.
Air pressure: 1100 hPa to 700 hPa
Altitude: 3000 m
Note:
The performance may be limited below 800 hPa or at altitudes above
2000m.