Accuracy of controls
146
Parameter
Range
Resolution
Accuracy
Trigger (inspiratory trigger)
(A)CV, P(A)CV:
Off, 0.5 to 15 L/min
V-SIMV, PSIMV, PS, CPAP:
0.5 to 15 L/min
0.5 L/min
N/A
Inspiratory Pressure Trigger (A)CV, P(A)CV:
Off, Very Low to Very High
PS, P-SIMV, V-SIMV, CPAP:
Very Low to Very High
N/A
Trigger sensitivity – Vented (S)T, P(A)C:
Very Low to Very High, Off
CPAP:
Very Low to Very High
N/A
Trigger sensitivity –
Mouthpiece
(A)CV, P(A)CV, PS
Off, Low to High, Touch
N/A
Rise Time
Min, 150 to 900 msec
50 ms
N/A
Flow Shape
100 (Constant), 75, 50, 25%
N/A
Ti Min
PS:
0.2 to 4.0 sec
(S)T, iVAPS:
0.1 to 4.0 sec
Adult: 0.1 sec
Paed: 0.05 sec
N/A
Ti Max
0.3 to 4.0 sec
Adult: 0.1 sec
Paed: 0.05 sec
N/A
Apnoea Interval
Adult: 15 to 60 sec
Paed: 5 to 30 sec
1 sec
±0.5 s
Apnoea Resp Rate
Adult: 4 to 50 bpm
Paed: 12 to 80 bpm
1 bpm
±2%
Apnoea Ti
When Volume Breath
option is set to Ti.
If Apnoea Response is
(A)CV + Alarm:
0.3 to 3 sec
If Apnoea Response is
P(A)CV + Alarm:
0.2 to 5 sec
Adult: 0.1 sec
Paed: 0.05 sec
±(20 ms + 5% of setting)
Apnoea Vt
If Apnoea Response is
(A)CV + Alarm:
Adult: 100 to 2500 mL
Paed: 50 to 300 mL**
Adult: 10 mL
Paed: 5 mL
Valved circuits:
±12 mL or 10%, whichever is greater
Apnoea flow shape
Constant
N/A