Troubleshooting
English 157
Alarm message
Action
High Vte
1.
Check the patient's status.
2.
Check double limb circuit adapter for contaminants or water.
3.
Inspect the double limb circuit adapter. If necessary, replace the circuit adapter.
4.
Verify that therapy and alarm settings are appropriate.
5.
Perform a Learn Circuit.
High Vti
1.
Check the patient's status.
2.
Inspect the circuit and expiratory module for leaks.
3.
Verify that therapy and alarm settings are appropriate.
4.
Perform a Learn Circuit.
Incorrect circuit adapter
1.
Check that the correct circuit adapter is installed for the selected circuit type.
2.
Perform a Learn Circuit.
Incorrect circuit attached
1.
Check that the circuit is correctly connected and matches the circuit type selected.
2.
Inspect the circuit, expiratory valve and proximal lines.
3.
Perform a Learn Circuit.
Internal battery degraded
1.
Connect the Astral to mains AC power.
2.
Return the device for service to replace the internal battery.
The internal battery run time indicator may no longer be accurate and should not be relied
upon.
Last self-test failed
1.
Perform a Learn Circuit.
2.
If problem persists, return the device for service.
Low internal battery
Connect the Astral to mains AC power to allow the battery to recharge.
Low FiO
2
1.
Check the patient's status.
2.
Check for leak.
3.
Check the oxygen supply and connections to the device.
4.
Verify that therapy and alarm settings are appropriate.
5.
Perform a Learn Circuit to recalibrate the oxygen sensor.
Low MVe
1.
Check the patient's status and airway.
2.
Inspect the circuit and the expiratory valve for obstruction or leaks.
3.
Verify that therapy and alarm settings are appropriate.
4.
Perform a Learn Circuit.
Low MVi
1.
Check the patient's status and airway.
2.
Inspect the circuit for obstruction.
3.
Verify that therapy and alarm settings are appropriate.
4.
Perform a Learn Circuit.
Low PEEP
1.
Check the patient's status.
2.
Verify that therapy and alarm settings are appropriate.
3.
Inspect the circuit and the expiratory valve for obstruction or leaks. When in use,
check for obstructions in proximal lines.
4.
Perform Learn Circuit.
Low pressure
1.
Check all circuit connections, especially the patient interface and the proximal sense
line.
2.
Verify that therapy and alarm settings are appropriate.
3.
Inspect the circuit and expiratory valve for damage or secretions.
4.
Perform a Learn Circuit.