Troubleshooting
English 159
Alarm message
Action
Safety system fault
1.
Check the patient's status.
2.
Transfer the patient to an alternate means of ventilation.
3.
Return the device for servicing.
System fault
1.
Check the patient's status.
2.
Perform a Learn circuit.
3.
If problem persists, or the device fails self-test, return the device for service.
Total power failure
1.
Check the patient's status and airway.
2.
Connect the device to AC mains.
3.
Check the battery charge level of the internal and external (if applicable) battery.
The total power failure alarm can only be silenced by connecting the device to
AC mains power.
Using internal battery
Confirm operation on internal battery is intended or restore external power.
If intending to use external power:
1.
Check the power cable connection between the mains or battery, the power supply
pack and the device.
2.
If using an external battery, check the external battery charge level and
replace/charge if empty.
3.
If using mains AC, check the supply output.
4.
If the problem continues, try an alternative external supply type (ie, Mains AC, Mains
DC or External Battery).
Ventilation stopped
Confirm it is appropriate to stop ventilation.