Appendix A: Definitions
168
Measured and calculated parameter definitions
The following measured and calculated parameters are displayed during configuration or during
ventilation. Each Ventilation mode details the parameters displayed.
Parameter
Definition
FiO
2
Average of percentage of Oxygen delivered to circuit.
I:E
I:E is the ratio of the inspiratory period to the expiratory period.
The measured I:E ratio is displayed as a monitored parameter during ventilation.
The expected I:E ratio is calculated and displayed on the settings screens if the Resp.
rate setting is not set to Off.
Leak
Leak is the average unintentional leak. It is reported as a percentage for Double limb
circuits and as a flow for Single limb circuits with intentional leak.
The measured Leak is displayed as a monitored parameter during ventilation.
MV
Minute Ventilation (MV) is the product of the Target Patient Rate (Target Pt Rate) and
expired tidal volume averaged over the last eight breaths.
The MV is displayed as a calculated parameter during iVAPS configuration.
MVe
Expiratory Minute Volume (MVe) is the product of the respiratory rate and expired tidal
volume averaged over the last eight breaths.
The measured MVe is displayed as a monitored parameter during ventilation.
MVi
Inspiratory Minute Volume (MVi) is the product of the respiratory rate and inspired tidal
volume averaged over the last eight breaths.
The measured MVi is displayed as a monitored parameter during ventilation.
Pressure
Pressure is the current airway pressure of the patient as measured at the patient port.
The measured Pressure is displayed as a monitored parameter during ventilation.
PEEP
End expiratory pressure (PEEP) is the airway pressure measured 50 ms prior to the end of
the last expiration.
The measured PEEP is displayed as a monitored parameter during ventilation.
Pmean
Mean airway pressure of the patient over the last breath.
% Spont cycle
% Spont cycle is the percentage of breaths that are spontaneously cycled over the past
20 breaths.
% Spont trig
% Spont trig is the percentage of breaths that are spontaneously triggered over the last
20 breaths.
The measured %Spont Trig is displayed as a monitored parameter during ventilation.
PIF
Peak Inspiratory Flow (PIF) is the maximum flow reached during the last inspiration.
The measured PIF is displayed as a monitored parameter during ventilation.
The expected PIF is calculated and displayed for volume controlled breaths on the
settings screens when the Inspiratory Phase Duration Option is set to Ti.
PIP
Peak Inspiratory Pressure (PIP) is the maximum airway pressure reached during the last
inspiration.
The measured PIP is displayed as a monitored parameter during ventilation.
Pulse rate
The measured Pulse rate (pulse) is displayed as a monitored parameter when a pulse
oximeter is used.