Appendix B: Ventilation parameters
English 171
Parameter
Valve Ventilation and Mouthpiece
Circuit
Leak Ventilation
Setting
AC
V
P(A
)C
V
PS
P-
SIM
V
V-
SIM
V
CP
AP
(S
)T
P(A
)C
CP
AP
iV
AP
S
PS
(cmH
2
O)
PS
•
Adult: 2 to 50 [7]
•
Paed: 2 to 50 [7]
Spontaneous breaths: V-SIMV,
P-SIMV
•
Adult: 2 to 50 [7]
•
Paed: 2 to 50 [7]
Min PS
(cmH
2
O)
0 to 50 [2]
Max PS
(cmH
2
O)
AutoEPAP Off: 0 to 50 [20]
AutoEPAP On: 8 to 50 [20]
IPAP
(cmH
2
O)
Adult: 4 to 50 [12]
Paed: 4 to 50 [12]
Vt
(mL)
V-SIMV (mandatory breaths),
(A)CV
•
Adult: 100 to 2,500 [500]
•
Paed: 50 to 300 [100]*
PIF
(L/min)
When Volume Breath option is
set to PIF:
(A)CV, V-SIMV (mandatory
breaths)
•
Adult: 10 to 120 [50]
•
Paed: 5 to 60 [10]
Ti
(sec)
When Volume Breath option is
set to Ti:
(A)CV, V-SIMV (mandatory
breaths)
•
Adult: 0.3 to 3.0 [1.0]
•
Paed: 0.3 to 3.0 [0.6]
P(A)CV, P-SIMV (mandatory
breaths)
•
Adult: 0.2 to 5.0 [1.0]
•
Paed: 0.2 to 5.0 [0.6]
P(A)C
•
Adult: 0.3 to 4.0 [1.0]
•
Paed: 0.3 to 4.0 [0.6]
Summary of Contents for Astral 100
Page 1: ...Clinical guide English ...