Appendix B: Ventilation parameters
176
Features
Valve Ventilation and Mouthpiece
Circuit
Leak Ventilation
Setting
AC
V
P(A
)C
V
PS
P-
SIM
V
V-
SIM
V
CP
AP
(S
)T
P(A
)C
CP
AP
iV
AP
S
Apnoea Vt
(mL)
If Apnoea Response is (A)CV +
Alarm:
Adult: 100 to 2500 [500]
Paed: 50 to 500 [100]
Apnoea Flow Shape
(%)
Constant
Apnoea PIF
(L/min)
When the Volume Breath is set
to PIF.
If Apnoea Response is (A)CV +
Alarm:
•
Adult: 10 to 120 [50]
•
Paed: 5 to 60 [10]
Apnoea P Control
(cm H
2
O)
If Apnoea Response is P(A)CV +
Alarm:
•
Adult: 2 to 50 [7]
•
Paed: 2 to 50 [7]
Safety Vt
(mL)
Adult: Off, 100 to 2500 [Off]
Paed: Off, 50 to 500 [Off]
Max PS
PS setting to 50 [PS +5]
P control max
(cm H
2
O)
P control to 50 [P c5]
IPAP Max
IPAP to 50 [IPAP +5]
*Adult Tapnoea can be extended to 15min when the mouthpiece interface is selected.