Using the Astral device
14
Using the Astral device
WARNING
Make sure the area around the device is dry, clean and clear of bedding or clothes or other objects
that could block the air inlet. Blocking the cooling vents could lead to overheating of the device.
Blocking the air inlet could lead to patient injury.
CAUTION
•
To prevent possible damage to the ventilator, always secure it to its stand or place it on a flat,
stable surface. For mobile situations, ensure the Astral device is contained within its
mobility bag.
•
Ensure the device is protected against water if used outdoors.
Connecting to mains power
To connect to mains power:
1.
Connect the DC plug of the supplied ResMed external power supply unit to the rear of the Astral
device.
2.
Connect the power cord to the ResMed power supply unit.
3.
Plug the other end of the power cord into the power outlet.
For further information on powering the Astral device, refer to Power management (see page 59).
Powering on the device
To power on the Astral device, simply press the green power on/off button at the back of the device. The
device will perform a system check as shown on the main screen.
On completion of the system check, the Patient Home screen and active program is displayed.
The Astral device is factory pre-set with one active program.
If more than one program displays on the Patient Home screen, the active program will be highlighted
orange. For further information, refer to Programs (see page 24).
Note:
Settings configured in the active program will be used when ventilation is started.