Using the Astral device
English
27
To adjust the Ventilation mode of the current active program:
1.
Access Clinical mode, the Main Settings screen will be displayed automatically.
2.
The current active Ventilation mode and settings are displayed.
3.
Press the
Ventilation mode
button. The
Select Ventilation Mode
screen will display, the current
active Ventilation mode will be highlighted orange. Select your required Ventilation mode.
4.
You will be returned to the Main Settings page. All changes you have made will be outlined in
orange, indicating a pending status until you press
Apply.
If the mode change is cancelled you will be
returned to the current active mode's primary settings page. Any modifications will be disregarded.
5.
Once settings have been applied the program is updated with the new Mode selection. Perform a
Learn Circuit.
Summary of Contents for Astral 100